DEN-1, Densitometer (suspension turbidity detector)
Product code : BSN.BS-050102-AAF
BSN.BS-050102-AAFDensitometer is designed for measurement of cell suspension's turbidity in the range of 0.3-5.0 McFarland units (100x10^6- 150x10^7 cells/ml).Densitometer provides the opportunity to measure solution turbidity in a …
Via de knop “offerte aanvraag” kan u vrijblijvend contact met ons opnemen voor verdere informatie.
DEN-1, Densitometer (suspension turbidity detector)
Product beschrijving
Densitometer is designed for measurement of cell suspension's turbidity in the range of 0.3-5.0 McFarland units (100x106 - 150x107 cells/ml).
Densitometer provides the opportunity to measure solution turbidity in a wider range (up to 15.0 McFarland units) however, it is necessary to remember that in this case the standard deviation values increase.
A densitometer is used for measurement of cell concentration (bacterial, yeast cells) during fermentation process, determination of microorganism sensitivity to antibiotics, microorganism identification using various test-systems, for measurement of absorption at the definite wavelength, as well as for quantitative estimation of concentration of colour solution, absorbing green light.
The operation principle is based on measurement of optical density with digital presentation of results in McFarland units. The unit is calibrated at the factory (for operation with 16 mm diameter glass tubes) and keeps calibration without power supply. However, if necessary it is possible to calibrate the unit by 2-6 points in 0.5-5.0 McFarland unit range. Both commercial standards (e.g. produced by BioMerieux, Lachema, etc.) and the cell suspensions prepared in a laboratory can be used for calibration. Possibility to restore factory calibration settings*
Following calibration kits are available on request:
- CKG16 for glass tubes with diameter 16 mm (latex particles);
- CKG18 for glass tubes with diameter 18 mm (polymer particles).
Two versions of the product are available:
- DEN-1 powered from external energy supply;
- DEN-1B powered both from external energy supply and from batteries (ÐÐ). Besides, DEN-1B operates with higher precision of measurements (up to 0,01 McF).
*NEW FUNCTION
DEN-1, Densitometer (suspension turbidity detector)
Product specificaties
Packaging Unit
1 Unit
Brand
Biosan
Accuracy, of the full scale
±3%
Display
LED
Display resolution
0.1 McF
External power supply
Input AC 100-240 V, 50/60 Hz; Output DC 12 V
Input current/power consumption (230V)
12 V, 80 mA / 1 W
Light source
LED
Measurement range
0.3-15.0 McF
Measurement time
1 s
Measurement wavelength (λ)
λ = 565 ±15 nm
Overall dimensions (WxDxH)
165 x 115 x 75 mm
Possibility to restore factory calibration settings
+
Sample volume
not less than 2 ml
Standard set
External power supply, A-16
Tube external diameter
16 mm (using A-16 adapter) or 18 mm (without adapter)
Weight
0.7 kg